Equipment

Angled Tube Head
Due to the anode heel effect, the x-ray beam is not uniform in the direction parallel to the anode-cathode axis of the x-ray tube. This property is used in mammography by aligning the cathode over the chest wall end (higher energy beam to image thicker area) and the anode over the nipple end (lower energy beam can penetrate thinner area).
C-Arm Design
The x-ray set is a c-arm. The whole gantry rotates so that the tube and breast table remain opposite each other.
Fixed Focus-Detector Distance (FDD)
The set is designed for a single examination and the focus-detector distance (FDD) or focus-to-film distance (FFD) of 65-66cm is considered optimum. This set FDD is a compromise between lower patient doses (lower doses with higher FFDs) and higher film doses (lower exposures with higher FFDs). Also, higher FDDs require longer exposures for a fixed mA resulting in more movement unsharpness.
Compression Device
The maximum force applied should be no greater than 200 N (approx. 20 kg weight). Standard compression forces are normally between 100 – 150 N. The compression plate is angled so that more of the breast is in contact with the compression paddle.
Fixed Field Size
Unlike in general radiography, only one type of examination is done meaning collimation creating fixed field sizes are all that are required.
Grids
Moving anti-scatter grids are used in normal mammography imaging. For magnification views, the breast support table is above the film to give magnification factors of around 1.8. In this case the large air gap between the breast and the film works to reduce scatter and so no grid is needed.
Automatic Exposure Control (AEC)
In screen-film mammography a separate AEC was required placed behind the cassette. With the currently used digital mammography the detectors act as the AEC. In screen-film radiography an AEC is required to ensure a suitable exposure to prevent under- or over-exposed film. In digital radiography, however, windowing can negate the effects of unsuitably exposed film and the AEC is more to ensure a suitable radiation dose for the patient and for the working parameters of the digital detector.
Target / filter material
- Need good differentiation of low contrast structures
- Need very high spatial resolution for micro-calcifications
Target
Need material that produces characteristic x-rays with energies of 17-20 keV (20-30 keV for larger breasts) to produce the best contrast. The commonly used material is Molybdenum (characteristic x-rays at 17.5 and 19.6 keV).
Filter
A filter with a k-edge of an energy just above the characteristic energies is used to remove the higher energy x-ray photons and make the beam as monoenergetic as possible. Molybdenum has a k-edge of 20 keV, just high enough so that the large increase in attenuation (k-edge) doesn’t fall into the characteristic energies produced at the molybdenum target.
Alternatives
Mostly MoMo (molybdenum target, molybdenum filter) but this does not give high enough energies for larger breasts.
- Rhodium has a k-edge at 23.3 keV and we can use a molybdenum target and rhodium filter (MoRh) to increase the amount of x-rays with energies in the range of 20 – 23.3 keV.
- Rhodium characteristic x-rays are at 20.2 – 22.7 keV. When used as a target this produces a beam with a mean energy that is higher than for MoMo and for MoRh.
- Tungsten (W) target and Rhodium filter. The x-ray output is reduced as no characteristic x-rays are produced (and, therefore, longer exposure times) but tungsten is much cheaper. It is mostly used in breasts with implants or that have been treated with radiotherapy as they are much larger and denser.
| Contrast | Radiation dose | |
| Highest | MoMo | MoMo |
| MoRh | MoRh | |
| RhRh | RhRh | |
| Lowest | WRh | WRh |
The mean energy of the spectrum decreases from WRh to MoMo. Lower energy photons have a higher probability of interacting with matter and, therefore, produces better contrast. However, the lower the energy, the greater the absorption, the more energy is deposited in the matter, and the higher the dose.
Summary
- General use: MoMo
- Dense breasts: MoRh or RhRh
Written by radiologists, for radiologists with plenty of easy-to-follow diagrams to explain complicated concepts. An excellent resource for radiology physics revision.
Spatial Resolution
A very high resolution is required to see microcalcifications. This is achieved via:
- Focal spot size
- Compression
- Anti-scatter grid
Small Focal Spot Sizes
Broad focal spot size = 0.3 mm
Fine focus focal spot size = 0.1 to 0.15 mm
From a point source, objects are easily resolved as separate on the film. However, with increasing focal spot size, the radiation comes from all parts of the source. The radiation creating the image does not provide a sharp image but has blurring at the edges. If the objects are too close together they can appear as one or an extra ‘object’ can be created.
Compression
Typical compression force is 100 – 150 N
The compression force:
- Lowers patient radiation dose as the attenuation of the compressed breast is lower and a lower exposure can be used
- Reduces scatter as the breast is less thick so there is less probability of scatter happening within the tissue
- Spreads the tissues out so that there is less overlaying of features
- Reduces geometric unsharpness by moving tissue closer to the image receptor
- Reduces movement unsharpness by holding the breast still
- The compressed breast is of more uniform attenuation
Anti-Scatter Grids
In mammography, moving grids are used for all contact (broad focus) images. For magnification images using a fine focal spot size or an air gap technique is used to reduce the amount of scattered radiation reaching the receptor meaning a grid is not required.
Altering Parameters
Parameters need to be altered to provide optimal imaging of different breasts. Two factors need to be taken into consideration:
- Thickness of breast
- Composition of breast
1. Thickness
In large breasts:
- More radiation absorbed – higher doses needed
- More scatter
- Increased beam hardening (lower contrast)
- Longer exposure needed at 28 kV MoMo, therefore, movement artefacts may occur
Thinnest breasts: MoMo at 25 kV
Thickest breasts: MoRh or even WRh for very thick breasts at 32 kV
2. Composition
With more dense breasts, higher doses are needed due to extra attenuation and more beam hardening. Due to beam hardening, the AEC may cut off the exposure prematurely (the measured exposure will be of a higher intensity). To ensure this doesn’t happen, one of two methods may be used:
- A pre-exposure determines whether the breast is as dense as expected for this thickness by looking at the dose rate and beam hardening.
- Adjustment on dose rate based on measuring the dose detected at the start of the examination and then adjusting the dose and exposure time as necessary.
Tomosynthesis
Superimposed tissue can mask pathology and, often, the pathology in breast disease can be very subtle. Breast tomography uses digital radiography to reconstruct planar images of sections of the breast. There are two main methods of acquiring breast tomosynthesis:
- The x-ray tube traverses along an arc acquiring images as it travels and the detector remains stationary
- The x-ray tube traverses along an arc and the detector also rotates
The images are then reconstructed using filtered back projection or iterative reconstruction (see Acquiring an image part 2).

Pros
- Provides enhanced lesion detection
- Reduces false positive recalls
- Allows more precise lesion localisation
Cons
- Higher radiation dose (approximately double)
- High contrast objects (e.g. surgical clips) can cause significant artefacts
- Longer interpretation time
- Requires substantially more data storage
Σ Summary
- Angled tube head
- Cathode over chest wall (thickest part of breast) to exploit anode heel effect (higher energy radiation at thickest part of breast)
- Focus-to-film distance
- Fixed at 65-66cm
- Target-filter material:
- General use – MoMo at 25 kV
- Dense breasts – MoRh or RhRh at 32 kV
- Compression
- Maximum pressure of 200N
- Lowers patient radiation dose
- Reduces scatter
- Spreads the tissues out so that there is less overlaying of features
- Reduces geometric unsharpness
- Reduces movement unsharpness
- More uniform attenuation
- Anti-scatter grids used
- Except in magnification view where air-gap used
- Focal spot
- Broad focal spot size = 0.3 mm
- Fine focus focal spot size = 0.1 to 0.15 mm
- Breast tomosynthesis
- X-ray tube travels in an arc +/- detector rotates
- Allows reconstruction of planar images
- Pros:
- Reduced recall rate
- Increased detection of pathology
- More precise lesion localisation
- Cons:
- Increased radiation dose (double)
- High contrast objects cause significant artefacts
- Longer interpretation times
- More data storage required
